The new TidalGuard HD is more compact and portable than ever, with a crisp high-definition color screen and the choice between mainstream and sidestream CO2 combined with digital pulse oximetry. It’s designed to provide immediate clinical feedback of actual expired CO2. The TidalGuard HD helps provide critical information during anesthesia, such as depth of anesthesia, cardiac output and perfusion, and reliable respiration.
Features:
General
Dimensions 4.8 x 2.9 x 0.9 in (125 x 73 x 23 mm)
Net weight 4.8 oz. (137 grams)
Electrical
External Power Supply: Includes AC adapter/charger
Power voltage: 100 – 240 V
Power frequency: 50/60 Hz
Battery (4xAA): NiMH 2400 mAh rechargeable or standard alkaline
Battery work time: SpO2 only - 10hr, CO2 + SpO2 – 4.5hr
Environmental
Operating Temperature: 32 – 122° F (0 – 50° C )
Operating Humidity: ≤95% relative (non-condensing)
Capnograph
Principle of Operation: Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts
Initialization Time: Capnogram within 15 seconds, full specifications within 2 minutes
Calibration: No regular calibration required
Capnostat® Mainstream CO2 Sensor
Shock Resistance: Withstands repeated 6 ft drops on tiled floor
CO2 (Carbon Dioxide)Range: 0 – 150 mmHgAccuracy: ±2 mm Hg (0 – 40 mmHg), ±5% of reading (41 – 70 mmHg), ±8% of reading (71 – 100 mmHg), ±10% of reading (101 – 150 mmHg) Resolution: 0.1 mmHg (0-69mmHg), 0.25 mmHg (70-150mmHg)Response Time: ≤60 ms
Respiratory Rate Accuracy: ±1 bpm Range0 – 150 bpm Resolution: 1 bpm
Pulse Oximeter
Principle of Operation: Red/Infrared absorption (Oxygen Saturation)
SpO2 Range: 0 – 100%Accuracy: ±2% SpO2 (for 70 – 100% SpO2) (1 SD), Unspecified for 0 – 69%Resolution: 1%
Pulse RateRange: 30 – 250 bpm Accuracy: 1 bpm or ±2%, whichever is greaterResolution: 1 bpm