Intuitive user interface
Quick to learn, easy to use
Preview and monitor
Real-time preview of 12-lead ECG and post-acquisition review of acquired ECG
Acquire and store
Internal storage for up to 60 digital ECG records
Interpretation algorithm
VERITASTM resting ECG interpretation algorithm with age (adult and pediatric) and gender criteria
Waveform display
Backlit 1/4 LGA LCD; 3-lead groups, 8 or 12-lead presentation
Data entry
Full keyboard with "one-touch" buttons for ECG acquisition, rhythm printing and ECG transmission
Reports
Full-size, (8.5" x 11" or A4) Z-fold paper output for multiple standard or Cabrera print options including 3+1, 3+3, 6-channel and 12-channel outputs (tray holds 250 sheets)
Connectivity options
Communication via serial port, internal modem, LAN or WAN
FAA capable
FAA/modem optional
Safety Parameters
Meets or exceeds the requirements of ANSI/AAMI ES1
Device Classification
Class I, Type CF defibrillation-proof applied parts
Physical Characteristics
Weight
11.25 lbs (5.1 kg) including battery (without paper)
Dimensions
15.5" x 17" x 4" (39.4 cm x 43.2 cm x 10.2 cm)
Power
Universal AC power supply (100-240 VAC at 50/60 Hz) 50 VA; internally rechargeable battery
ECG Acquisition
Input impedance
Meets or exceeds the requirements of ANSI/AAMI EC11
Standard leads
I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Frequency response
0.05 to 300 Hz
Filters
High-performance baseline filter; AC interference 50/60 Hz; low-pass filters 40Hz, 150 Hz or 300 Hz
A/D conversion
20 bits
Digital sampling rate
10,000 samples/sec/channel used for pacemaker spike detection
Analysis rate
1,000 samples/sec/channel used for recording and analysis
Thermal printer
Computer-controlled dot array; 8 dots/mm
Thermal printer speeds
5, 10, 25 or 50 mm/s
Gain settings
5, 10, or 20 mm/mV
Report print formats
Standard or Cabrera: 3+1, 3+3 channel, 6-channel, 12-channel
Rhythm print formats
3-channel, 6-channel or 12-channel with configurable lead groups